Integrated Publishing
Army Medical Portable VentilatorPortable VentilatorDevelopmentTable of ContentsIntroductionCredit Awarded-cont.Lesson AssignmentLesson 1: Operation and General Maintenance ProceduresFigure 1-1. Babybird 5900 pediatric ventilator.Major ComponentsDelivery Circuit.Figure 1-4. Flow-bypass valve.Figure 1-5. Mark 2 servo.Inspiratory time-limit control. Compound-lockout cartridge.Breathing Circuit.Inspiratory tube.Overpressure governor.Section II. Theory of OperationsSpontaneous Breathing ModeFigure 1-12. Inhalation phase of the spontaneous mode.Figure 1-13. Expiratory path in any mode of operation.Controlled Intermittent Mandatory Ventilation ModeFigure 1-14. Inhalation phase of the controlled IMV mode.Controlled Intermittent Mandatory Ventilation Mode-cont.Figure 1-15. Exhalation phase with negative pressure in the controlled IMV mode.Section III. Preventive Maintenance Checks and ServicesCorrect Minor Deficiencies After Reassembly.Exercises, Lesson 1Exercises, Lesson 1-cont.Exercises, Lesson 1-cont.Exercises, Lesson 1-cont.Solutions to Exercises, Lesson 1Lesson AssignmentLesson 2: Locating and Replacing Defective ComponentsCalibrate the VentilatorPrepare the Ventilator for Calibration.Figure 2-1. Ventilator mounted for calibration with back and side panel removed.Calibrate the Ventilator.Table 2-1. Source pressure gauge troubleshooting.Table 2-2. Airway manometer troubleshooting.Table 2-3. Flow test device pressures.Nebulization system.Table 2-4. Flow system troubleshooting-cont.Table 2-4. Flow system troubleshooting-concludedExpiratory-flow gradient.Table 2-5. Nebulization system troubleshooting.Outflow valve continuous positive airway pressure control (CPAP).Table 2-6. Expiratory flow gradient troubleshooting-concludedPressure relief valve.Mark-2 timing system.Table 2-9. Mark-2 timing system troubleshooting.Inspiratory lockout (compound-lockout cartridge).Table 2-10. Inspiratory relief pressure troubleshooting.Inspiratory time limit.Table 2-11. Inspiratory lockout troubleshooting.Figure 2-8. Control knob set at 3 seconds.Inspiratory time (Mark 2).Table 2-13. Inspiratory time (Mark 2) troubleshooting.Low pressure alarm system.Calibration Test.Flow system.Pressure relief valve.Section II. Troubleshooting and Removal and Replacement ProceduresTable 2-16. Troubleshooting at start up in either mode-cont.Table 2-16. Troubleshooting at start up in either mode-cont.Table 2-17. Troubleshooting in either mode of operation-cont.Table 2-17. Troubleshooting in either mode of operation-cont.Table 2-17. Troubleshooting in either mode of operation-concludedTable 2-18. Troubleshooting in the spontaneous breathing mode.Remove and Replace Major ComponetsRemove and Replace the PAP Gauge (Part Number 9407R). Remove and Replace the Bypass Valve.Remove and Replace the Nebulizer Valve.Exercises, Lesson 2Exercises, Lesson 2-cont.Exercises, Lesson 2-cont.Solutions to Exercises, Lesson 2
Army Medical
Portable VentilatorPortable VentilatorDevelopmentTable of ContentsIntroductionCredit Awarded-cont.Lesson AssignmentLesson 1: Operation and General Maintenance ProceduresFigure 1-1. Babybird 5900 pediatric ventilator.Major ComponentsDelivery Circuit.Figure 1-4. Flow-bypass valve.Figure 1-5. Mark 2 servo.Inspiratory time-limit control. Compound-lockout cartridge.Breathing Circuit.Inspiratory tube.Overpressure governor.Section II. Theory of OperationsSpontaneous Breathing ModeFigure 1-12. Inhalation phase of the spontaneous mode.Figure 1-13. Expiratory path in any mode of operation.Controlled Intermittent Mandatory Ventilation ModeFigure 1-14. Inhalation phase of the controlled IMV mode.Controlled Intermittent Mandatory Ventilation Mode-cont.Figure 1-15. Exhalation phase with negative pressure in the controlled IMV mode.Section III. Preventive Maintenance Checks and ServicesCorrect Minor Deficiencies After Reassembly.Exercises, Lesson 1Exercises, Lesson 1-cont.Exercises, Lesson 1-cont.Exercises, Lesson 1-cont.Solutions to Exercises, Lesson 1Lesson AssignmentLesson 2: Locating and Replacing Defective ComponentsCalibrate the VentilatorPrepare the Ventilator for Calibration.Figure 2-1. Ventilator mounted for calibration with back and side panel removed.Calibrate the Ventilator.Table 2-1. Source pressure gauge troubleshooting.Table 2-2. Airway manometer troubleshooting.Table 2-3. Flow test device pressures.Nebulization system.Table 2-4. Flow system troubleshooting-cont.Table 2-4. Flow system troubleshooting-concludedExpiratory-flow gradient.Table 2-5. Nebulization system troubleshooting.Outflow valve continuous positive airway pressure control (CPAP).Table 2-6. Expiratory flow gradient troubleshooting-concludedPressure relief valve.Mark-2 timing system.Table 2-9. Mark-2 timing system troubleshooting.Inspiratory lockout (compound-lockout cartridge).Table 2-10. Inspiratory relief pressure troubleshooting.Inspiratory time limit.Table 2-11. Inspiratory lockout troubleshooting.Figure 2-8. Control knob set at 3 seconds.Inspiratory time (Mark 2).Table 2-13. Inspiratory time (Mark 2) troubleshooting.Low pressure alarm system.Calibration Test.Flow system.Pressure relief valve.Section II. Troubleshooting and Removal and Replacement ProceduresTable 2-16. Troubleshooting at start up in either mode-cont.Table 2-16. Troubleshooting at start up in either mode-cont.Table 2-17. Troubleshooting in either mode of operation-cont.Table 2-17. Troubleshooting in either mode of operation-cont.Table 2-17. Troubleshooting in either mode of operation-concludedTable 2-18. Troubleshooting in the spontaneous breathing mode.Remove and Replace Major ComponetsRemove and Replace the PAP Gauge (Part Number 9407R). Remove and Replace the Bypass Valve.Remove and Replace the Nebulizer Valve.Exercises, Lesson 2Exercises, Lesson 2-cont.Exercises, Lesson 2-cont.Solutions to Exercises, Lesson 2
Privacy Statement - Copyright Information. - Contact Us