18.
To test that the temperature controller maintains the proper temperatures, you
observe DS1 and DS2 blinking at what ratio when there is no film in the
AFP 14-X3 X-ray Film Processor?
a. 50:50
b. 20:80
c.
80:20
d. 30:70
19.
If the AFP 14-X3 X-ray Film Processor's dryer heaters exceed approximately
165F the dryer safety thermostat(s) cuts off power. You must do which of the
following?
a. Remove the front dryer and then the lift out dryer, in order to test
independently of one another.
b. Use the automatic reset button and test a second time.
c.
Allow the dryers to cool before pressing the manual reset button.
d. Remove the dryer assembly and clean it on the first sign of a problem.
20.
Which of the following do you use or observe when checking the replenishment
system on the AFP 14-X3 X-ray Film Processor?
a. In a low volume mode, the pump should run only when you press the film
switch.
b. In a high volume mode you press the film switch and observe the pump to see
that it runs and that when you release the switch, it stops.
c.
In a high volume mode you need to check that the pump is activating during
the jog cycle.
d. In both modes, you should observe the led DS3 on the J3 PC board, which is
the pump run signal.
Check Your Answers on Next Page
MD0359
1-44