PROBABLE CAUSE
CORRECTIVE ACTION
SYMPTOM
19. Monitor temperature
Battery condition low.
Replace the battery.
led safe lamp on; power
on led lamp on; Power
failure led lamp flashing;
audible alarm pulsing.
20. Monitor Temperature
Monitor temperature led
Replace lamp(s).
led lamps off; power on led
lamp(s) defective.
lamp on; power failure led
lamp off; audible alarm off.
Circuit defective.
Determine reason and
correct.
21. Monitor temperature
Circuit(s) defective.
Determine reason and
led lamps on; power on led
correct.
lamp on; power failure led
lamp off; audible alarm off.
22. Monitor temperature
Battery test switch in off
Place in up position.
led safe lamp off; power
(down) position.
on led lamp off; power
failure led lamp flashing;
audible alarm pulsing.
End of Appendix
MD0367
A-8