Cat. No. 0287
Cat. No. 0269 & 0271
12-place 15ml horizontal rotor with 12-0901
4-place horizontal rotor with 4 multiple
shields.
carriers for 36 10x75mm tubes or
28 12x75mm tubes (0920 & 0921).
TIP RADIUS-6.37 inches
TIP RADIUS-5.0 inches
MAX SPEED-2610rpm
MAX SPEED-3600rpm
RCF-1232
RCF-1840
Cat. No. 0289
Cat. No. 0293
24-place 15ml angle rotor with 24-0902
24-place angle rotor with 24-0904 shields.
shields.
TIP RADIUS-5.75 inches
TIP RADIUS-5.0 inches
MAX SPEED-2760rpm
MAX SPEED-3400rpm
RCF-1244
RCF-1641
Cat. No. 0291
24-place 15ml angle rotor with 24-0901
shields.
TIP RADIUS-6.37 inches
MAX SPEED-2500rpm
RCF-1131
Figure 1-2. Maximum angular velocities and relative centrifugal Forces for Dynac II
Centrifuge rotors (Model 0103 at 120 volts alternating current (vac),
60 hertz (Hz) and Model 0106 at 220vac, 50 Hz).
(Concluded)
d. Timer. Spinning time is controlled by a 30-minute timer with a hold position for
continuous operation. Operating the timer automatically turns on the centrifuge, provided
the power switch is on and the cover is securely latched.
(1) To set the timer, turn the knob clockwise to the desired setting. The
centrifuge will now operate to the end of the timed cycle and then shut off automatically.
(2) For continuous operation, turn the knob clockwise, to the HOLD position.
The centrifuge will now operate continuously until it is manually shut off. To shut off, turn
the knob counterclockwise to "0."