Figure 1-2. Major components.
f. Loosen the clamp knobs beneath the front end of the back section. Guide the
head extender rods into the bearing holes at the front end of the back section. Slide the
head section rods all the way in. Hand tighten the clamp knobs.
g. Remove the mattress pads from the storage case. Turn the pads over so the
elastic bands are facing up. Carefully remove the Formica x-ray tops from each table
top section. Lay the tops face down (mount pins up). Pull the elastic straps up and
over each pin.
CAUTION:
USE EXTREME CAUTION when removing the x-rays tops. They are
made of Formica and can be easily broken.
h. Remove the remaining parts from the storage case and install them on the
side rails of the table top.
1-4.
PERFORM OPERATIONAL CHECK-OUT PROCEDURES
To ensure that the table operates properly, perform the following procedures.
a. Visually inspect the condition of the operating table for loose, broken, or
missing parts.
MD0370
1-4