g. Select tomographic technique factors:
(1)
60 kVp.
(2)
50 mAs, approximately.
h. Select the cut level of 12.5 cm (see figure 6-3).
Figure 6-3. Select the cut level of 12.5 cm.
i. Make the exposure.
j.
Process the test film.
k. Record the technique factors used and the tomographic parameters on the
data form and the film.
l.
Repeat the procedure for other exposure angles and tomographic motion.
MD0062
6-6