6.
Which of the following responses correctly identifies an action you will take to
install new cabling into the scissors arms?
a. Connect the cabling single lines to multi-pin numbers 1, 2, 3, 4, 5.
b. Connect the cabling single lines to multi-pin numbers 2, 3, 4, 5.
c.
Connect the cabling single lines to multi-pin numbers 1, 2, 3, 4.
d. Connect the cabling single lines to multi-pin numbers 3, 4, 5, 6.
7.
To inspect the rotational parts, you remove the X-ray head from the scissors arm.
Which of the following is a step you use?
a. Disconnect wires in the U-channel but do not disconnect the ground wire.
b. Before you actually remove the X-ray head, check the rotational stop
assembly.
c.
When reassembling, grease the bracket and secure the shaft with the bracket
to the arm.
d. When reassembled, verify you have rotation beyond 1 1/2 turns.
8.
Which of the following statements is true about exchanging the pcbs?
a. You may exchange pcb D1 only with a soldering tool.
b. You may exchange pcb D3 only with a soldering tool.
c.
You must use a soldering tool to exchange pcb D5.
d. All pcb are factory calibrated and cannot be adjusted.
9.
Which of the following is a procedure you use to exchange the carrying case
gasket?
a. Remove the old gasket, clean, and reinstall.
b. Clean the gasket retainer groove with a scraper.
c.
Use new adhesive to repair cuts in the gasket.
d. Do not repair gaskets; replace them.
Check Your Answers on Next Page
MD0361
4-21