1-21. IMMERSION OF HANGERS IN SOLUTIONS
When placing hangers loaded with film in solutions, dip them quickly and
carefully. After complete immersion, raise and lower the films about two inches several
times to remove air bells and completely bathe the surfaces of the film. Make sure
hangers are separated by about three-fourths inch. Do not remove them entirely from
the solution until you are ready for the next operation. By keeping the level of the
solutions one-inch below the tops of the tanks, the crossbars of the hangers are not
covered by the developer of fixer solutions. However, the crossbars should be
immersed in the wash water.
MD0954