Coded Troubleshooting Guide (continued)
CAUTIONS (continued)
CODE/MALFUNCTION
POSSIBLE CAUSE
REMEDY
04
1. Exposure temperature
1. Reset exposure
Positive pulse retarded
setting incorrect.
temperature.
2. Turn on steam supply.
(exceeds 5 minutes)
2. Steam supply shut off.
3. Steam pressure
3. Check and ensure steam
supply pressure is 40-50psig
inadequate.
dynamic.
4. Clean or repair.
not closed.
5. Clean or repair.
not opening properly.
6. Solenoid valve 8SOL
6. Clean or repair.
not opening properly.
7. Clean strainer.
7. Steam supply strainer
plugged.
8. Pressure release
8. Repair jacket trap.
through jacket trap
excessive.
9. Pressure release
9. Repair chamber trap.
through chamber trap
excessive.
06 (16x16)
1. Exposure temperature
1. Reset exposure
Temperature retarded
setting incorrect.
temperature.
exceeds the following
2. Steam supply shut off.
2. Turn on steam supply.
times:
3. Steam pressure
3. Check and ensure steam
Wrapped goods
inadequate.
supply pressure is 40-50psig
4 minutes
dynamic.
Unwrapped instruments -
4. Clean or repair.
6 minutes
not closing properly.
Liquids - 25 minutes
5. Clean or repair.
not opening properly.
6. Solenoid valve 8SOL
6. Clean or repair.
not opening properly.
7. Steam supply strainer
7. Clean strainer.
plugged.
8. Pressure release
8. Repair chamber trap.
through chamber trap
excessive.
MD0354
A-2