Coded Troubleshooting Guide (continued)
CAUTIONS (CONTINUED)
CODE/MALFUNCTION
POSSIBLE CAUSE
REMEDY
06 (20x20)/
1. Exposure temperature 1. Reset exposure
Temperature retarded
setting incorrect.
temperature.
exceeds the following
2. Steam supply shut off. 2. Turn on steam supply.
times:
3. Steam pressure
3. Check and ensure steam
Wrapped goods - 4
inadequate.
supply pressure is 40-50psig
minutes
dynamic.
Unwrapped instruments -
4. Clean or repair.
9 minutes
not closing properly.
Liquids - 30 minutes
5. Clean or repair.
not opening properly.
6. Solenoid valve 8SOL
6. Clean or repair.
not opening properly.
7. Steam supply strainer 7. Clean strainer.
plugged.
8. Pressure release
8 Repair chamber trap.
through chamber trap
excessive.
10
1. Steam supply shut off. 1. Turn on steam supply.
Exposure temperature
2. Steam pressure
2. Check and ensure steam
reset (set point -2OC)
inadequate.
supply pressure is
40-50psig dynamic.
3. Clean or repair.
not closing properly.
4. Clean or repair.
not closing properly.
5. Solenoid valve 8SOL
5. Clean or repair.
not closing.
6. Chamber drain screen 6. Clean screen.
clogged.
14
1. Exposure temperature 1. Reset exposure
Excessive exposure
setting incorrect.
temperature.
temperature (exceeds
2. Clean or repair.
6OC over set point)
not closing properly.
3. Solenoid valve 8SOL
3. Clean or repair.
not closing properly.
18
1. Air-in filter.
1. Replace filter.
Air-in retarded (exceeds 2
2. Solenoid valve 3SOL
2. Clean or repair.
minutes)
not opening properly.
MD0354
A-3