│ Problem
│ Potential Cause │ Corrective Action
│
╞════════════════╪══════════════════╪═══════════════════════════╡
│ Cannot adjust │ Leak in the
│ 1. Apply a liquid leak
│
│ compound-
│ inspiratory
│ detector to both ends of │
│ lockout
│ tubing system.
│ tube connected to the
│
│ cartridge for │
│ compound-lockout cartridge│
│ inspiratory
│
│ and correct any connection│
│ lockout at
│
│ which bubbles.
│
│ 12 to 14psig. │
│
│
│
│
OR
│
│
│
│
│
│
├────────────────┼──────────────────┼───────────────────────────┤
│
│ Internal
│ 2. Remove cartridge from │
│ malfunction of
│ the ventilator and
│
│
│
│ compound-lockout │ disassemble it. Replace │
│
│ cartridge (P/N
│ worn component parts.
│
│
│ 5913).
│
OR
│
│
│
│ 3. Install a new compound-│
│
│
│ lockout cartridge.
│
├────────────────┼──────────────────┼───────────────────────────┤
│ Audible alarm │ Alarm assembly
│ 4. Ensure the alarm is
│
│ not activated │ not secured to
│ firmly attached to the
│
│ after an
│ bottom of
│ ventilator bottom.
│
│ inspiratory
│ ventilator.
│
│
│ lockout.
│
OR
│
│
│
├──────────────────┼───────────────────────────┤
│
│ Faulty reed (P/N │ 5. Replace reed.
│
│
│ 1866) in alarm. │
OR
│
│
│
│ 6. Replace with new alarm │
│
│
│ (P/N 5127).
│
│
│
OR
│
│
│
├──────────────────┼───────────────────────────┤
│
│ Malfunction in
│ 7. See actions 1 through │
│
│ compound-lockout │ 3 above.
│
│
│ cartridge.
│
│
└────────────────┴──────────────────┴───────────────────────────┘
Table 2-11. Inspiratory lockout troubleshooting.
(e) Slowly rotate the inspiratory time-limit stem counterclockwise until
an inspiratory lockout occurs at 3 seconds (plus or minus 0.2 second). Begin your timing
at the start of an inspiratory cycle and stop timing when the unit locks out, not at the
sound of the alarm.
(f) Confirm a 3 second lockout (plus or minus 0.2 second) and then
mark the end of the inspiratory time-limit stem at 12 o'clock with a pencil.
(g) Push the reset button and rotate the inspiratory time-limit stem
clockwise for a 5.5- to 6.5-second lockout time.