Figure 6-8. Tomographic wedge parallel to the direction of the tube travel.
h. Set the control panel:
(1)
60 kVp.
(2)
30 mAs approximately.
i.
Make the tomographic exposure.
j.
Process the test film.
k. Repeat the test for other tomographic motion sweeps.
l.
Process the films.
6-22. EVALUATION AND INTERPRETATION
a. Ensure that the radiographs and the recording form include the following
information:
(1)
Tomographic motion.
(2)
Sweep speed.
(3)
Exposure angle.
(4)
Cut level.
MD0062
6-15