Coded Troubleshooting Guide (concluded)
CAUTIONS (CONTINUED)
CODE/MALFUNCTION
POSSIBLE CAUSE
REMEDY
20
1. Clean or repair.
Pressure in chamber
not opening.
(Exceeds 2psi over or
2. Clean or repair.
under atmospheric
not closing properly.
reading)
3. Solenoid valve 6SOL
3. Clean or repair.
not opening properly.
4. If problem cycle was
4. Clean bleed
"Liquids," the chamber
valve.
bleed valve may be
clogged.
22
Vacuum leak test. In progress.
End of Appendix A