k. Controls OFF Switch (L). Use this switch to turn off the control/display
indicators, selectors, and switches. The maintenance of jacket temperatures ceases.
Times and temperatures for the last selected program for all programmed cycles remain
in the memory. They do not need to be reset when sterilizer operation is restarted
unless new times and temperatures are desired for that cycle.
NOTE:
When using the controls off switch to abort a cycle in response to a coded
warning or caution, be sure to record the code for future reference. Warning
or caution codes are erased when the control panel is turned off unless
equipped with a data logger.
l. Unseal Door Switch (M). Single door units do not require the use of this
switch for normal operation because the door unseals automatically. Pressing an
unsealed door switch closes the related three-way door gasket solenoid to steam
pressure and opens it to drain, which releases the door-to-chamber seal. An unseal
door switch does not function on a double door unit unless the opposite door is sealed.
m. Seal Door Switch (N). Single door units require the use of this switch for
normal operation. Press a seal door switch to open the related three-way door gasket
solenoid to steam pressure and close it to drain, creating a door-to-chamber seal.
n. Cycle Start Switch (P). Use this switch to start an automatic cycle on single
door units. On double door units, it starts an automatic cycle after both doors are
sealed.
o. Wrapped Goods Selector (Q). The wrapped goods selector permits the
energizing of predetermined components that will accomplish a wrapped goods cycle.
When more than one of the same type of cycles are run in a series, press the selector
for the first cycle only.
p. Unwrapped Instruments Selector (R). Press this selector to permit the
energizing of predetermined components to accomplish an unwrapped instruments
cycle. When more than one of the same cycle are run in a series, press the selector for
the first cycle only.
q. Liquids Selector (S). Press this selector to permit the energizing of
predetermined components to accomplish a liquids cycle. When more than one of the
same cycle are run in a series, press the selector for the first cycle only.
MD0354