12.
Which of the following statements is true when checking for correct chain tension
on the AFP 14-X3 X-ray Film Processor?
a. The processor drive chain is located in the crossover assembly.
b. Overly taunt chain tension causes excessive sprocket wear.
c.
The drive chain tension is set at the manufacturer's and should not require
adjustment.
d. To remove the chain, you remove the connecting link from the drive chain,
then remove the chain from the sprockets.
13.
Which of the following statements is true when checking for circulation of the tank
solutions with the AFP 14-X3 X-ray Film Processor?
a. The developer and fix circulation pumps are identical.
b. You must lubricate the motors at least quarterly.
c.
The magnetically driven pump impellers are directly attached to the pump
motor drive shafts.
d. To check for a stuck impeller, run the pumps when they are dry to see if you
get excessive heat.
14.
Which of the following is a procedure for checking the air system for the
AFP 14-X3 X-ray Film Processor?
a. Call the manufacturer's representative for consultation when the blower
motors remain off during the developer solution warm-up.
b. Check power to the blower motor while the blower is running.
c.
When you check the amp draw between TS3-4 and TS3-7 there should be no
line voltage.
d. If the blower is not working at all, check fuse F2.
MD0359
1-42