Figure 1-7. Wet section rack.
d. Perform Quarterly Preventive Maintenance Checks Services. After
performing the daily, weekly, and monthly PMCS and after it has been three months
since the previous quarterly PMCS, the operator performs the quarterly PMCS. It
consists of adding several drops of SAE-30 oil to the output shaft bearing on the drive
motor reducer. Refer to figure 1-3.
Figure 1-8. Dryer rack.
e. Perform Semi-Annual Preventive Maintenance Checks Services. After it
has been six months from the previous semi-annual PMCS, the operator performs
semi-annual PMCS. The semi-annual PMCS consists of adding a moderate amount of
AFP-8827 grease to the drive shaft thrust bearing. Refer to figure 1-4.
MD0359
1-12