(6)
Move the collimator horizontally to each of the available SIDs (40 or 72
inches).
(a) An SID indicator should switch ON.
(b)
The collimator field-size indicators (front panel) should rapidly
adjust and stop.
(c)
The HOLD indicator should switch OFF.
(d) The READY indicators should switch ON.
(e) The light field size should be about the same size as the cassette in
the wall cassette holder.
(7) Remove the cassette from the table cassette tray. The removal of the
cassette from the table cassette tray should not produce a response.
k. Checkout for Tilting Tables.
(1) Tilt the table to full upright. At +79 degrees, and without a cassette in
the table Bucky, the following should occur.
(a)
Light-emitting diode-2 on the logic PCB should switch ON.
(b) The READY indicators should switch OFF.
(c)
The MANUAL indicator should switch ON.
(2) Angle the collimator to the +90 degrees beam-left position. Insert a
cassette into the cassette tray and fully insert the tray into the table Bucky.
(a)
Light-emitting diode-2 on the logic PCB should switch OFF.
(b) The HOLD indicator should switch ON.
(c)
The MANUAL indicator should switch OFF.
(3) Move the collimator horizontally to each of the available SID's to the
tilted table (40 or 72 inches).
(a) An SID indicator should switch ON.
(b)
The collimator field-size indicators (front panel) should rapidly
adjust and stop.