(b) Vary the SID of the SFD while observing the fluoroscopic image.
Verify the image size remains the same as in paragraph (3)(a) above and all collimator
blades are visible.
(c) Close the collimator shutter blades to their minimum using the
shutter controls on the SFD. Verify a small square or slightly rectangular image is
displayed. If it is not, adjust "COLL MIN" potentiometer, R22, on the SFD interface
board until the desired image is obtained.
(4) One-on-one (full) film adjustment.
(a) Lower the SFD to its minimum SID. Select the radiographic mode
on the generator. Load an unexposed cassette into the SFD and place the film in the
"park" position. Open the shutter blades to their maximum open position and select the
one on one format.
(b) Make an exposure. Make a radiographic exposure and develop the
film. Measure the exposed area and verify it is between 9.25 inches, x 9.25 inches, and
9.0 inches x 9.0 inches in size. If it is not, adjust the "FULL" potentiometer, R14, on the
SFD interface board to increase or decrease the exposed area as required.
(c) Repeat the procedure if necessary. Repeat this procedure until the
exposed area is within the limits described in paragraph (4) (b) immediately above.
Two-on-one longitudinal film adjustment.
(5)
(a) With the SFD at minimum SID and the radiographic mode selected
on the generator, insert an unexposed film cassette and advance it to the "park"
position. The collimator blades should be set to their maximum open position.
(b) Select the two-on-one longitudinal format and make two radiographic
exposures. Develop the film and measure the exposed areas. The dimensions of each
exposure should be between 9.25 inches x 4.625 inches and 9.0 inches x 4.5 inches
with no more than 3/16-inch gap or overlap between the images. If it is not, adjust
"HALF" potentiometer, R15, on the SFD interface board to increase or decrease the
exposed area as required.
(c) Repeat this procedure until the exposed areas are within the limits
described in paragraph (5) (b) immediately above.
(6)
Two-on-one transverse film adjustment.
(a) With the SFD at minimum SID and the radiographic mode selected
on the generator, insert an unexposed film cassette and advance it to the "park"
position. The collimator shutters should be set to their maximum open position.