(2) Remove the cabling to the generator. Refer to figure 4-2 and use the
following steps.
(a) Loosen the safety strap (37.1) until you have opened the scissors
arm (37) about 7 inches.
(b) Secure the safety strap with adhesive tape.
(c)
Remove the cover cap (93).
(d) Unscrew the cover (71) and cover cap (72).
Figure 4-2. Scissors arm cabling fitting, part one.