(e) Remove the scissors arm cabling (127.1) from the generator (38)
by gently pulling the prongs of plug connections 3, 4, 5, and 6 from the generator.
(f)
Unscrew the grounding screw on the arm to disconnect the
grounding wire.
(g) Unscrew and remove the angle support (bracket) (95).
(h)
Pull out the Heliodent generator (38).
(3) Prepare to remove cabling from scissors arm number one. Refer to figure 4-3 for
the following steps.
(a) Unscrew the axle (21) (4 screws) and remove it from scissors arm.
(b)
Remove the cable clip (A).
Figure 4-3. Scissors cabling arm fitting, part two.
MD0361
4-5