3.
Which of the following procedures do you use to check the incoming water supply
on the X-OMAT?
a. Open the wash tank drain valve and fully turn on both mixing valves.
b. Verify the hot water temperature in the wash tank is over 15F (8.3C) below
the temperature set at the mixing valve.
c.
Verify the cold water temperature in the wash tank is 15F (8.3C) below the
temperature set at the mixing valve.
d. Verify that the time necessary to fill the wash tank to overflowing is
approximately 60 seconds.
4.
Which of the following procedures do you use to calibrate the mixing valve on the
X-OMAT?
a. Turn the mixing valve handle screw to the cold position to adjust the water
temperature.
b. Turn the adjusting screw clockwise to reduce the outlet temperature.
c.
DO NOT turn the screw more than one full turn at a time.
d. If you cannot maintain proper temperature check the temperature of the hot
water supplying the mixing valve.
5.
Which of the following procedures do you use to calibrate the dryer thermostat on
the X-OMAT?
a. Place a thermometer of known accuracy in the bottom of the dryer to measure
actual temperature.
b. Place a thermometer of known accuracy in the dryer with the bulb resting
against the dryer surface.
c.
Reset the temperature dial if there is a variance of more than 2F between
the thermostat setting and thermometer reading.
d. Reset the temperature dial if there is a variance of more than 5F
between the thermostat setting and thermometer reading.
MD0359
3-10