(3) Disconnect the external power source and disconnect the chair power
cord from the receptacle in the lift.
(4) Remove the four nuts, lock washers and the mounting screws that
secure the chair to the lift. Remove the chair.
(5)
Remove the four mounting screws that secure the base cover. Refer to
figure 3-3.
(6)
Loosen the nut that secures the rotation lock pedal.
(7)
Lift the main cover off over the top of the lift.
Be sure you have disconnected the external power cord.
CAUTION:
(8)
Remove the gear motor and the capacitor.
(a) Remove the junction box cover, disconnect the motor wires and
remove the motor leads from the box. Refer to figure 3-5.
Figure 3-5. PLR-200 wiring diagram.
MD0373
3-7