(c)
If no action is required, place an "X" or "√" in the "ok" column.
(d) If action is required, annotate the action needed in the "action-
needed" column.
(e) When the action has been taken, enter the date and your initials in
the "action-taken" column.
(7)
Block 13, internal discharge of stored energy setting. Refer to figure 2-9.
(a) Charge the defibrillator to its maximum energy level.
(b) Test and verify the discharge of stored energy through the internal-
(c)
If no action is required, place an "X" or "√" in the "ok" column.
(d) If action is required, annotate the action needed in the "action-
needed" column.
(e) When the action has been taken, enter the date and your initials in
the "action-taken" column.
(8)
Block 14, energy delivered after one minute. Refer to figure 2-9.
(a)
Set the defibrillator control setting to the maximum setting and
charge.
(b) Measure and record the maximum energy delivered after a one
minute storage.
(c)
If no action is required, place an "X" or "√" in the "ok" column.
(d) If action is required, annotate the action needed in the "action-
needed" column.
(e) When the action has been taken, enter the date and your initials in
the "action-taken" column.
Block 15, output of tenth repeated discharge. Refer to figure 2-9.
(9)
(a) Charge and discharge the defibrillator at maximum setting 10 times.
(b) Measure and record the energy output of the 10th repeated
discharge.
MD0358
2-13