m. Table Top Movement Right-Limit Adjustment.
(1)
Verify adjustment is necessary.
(a) Place a piece of tape on the table top frame 30 inches from the
right (foot) end. Mark exactly the 30-inch position for reference.
(b) Press the top-right switch until the table top stops. If the table top
stops when the 30-inch reference mark is aligned with the foot-end of the table body, no
adjustment is necessary.
WARNING
Under no circumstances is any adjustment to be performed while the
table top is being driven. Serious injury could result. Make adjustments
to cam orientation first, and then alternated with verification through
actuating the table top drive system.
(2)
Remove the table top and the table sub-top.
(3) Locate the top-drive control-switch assembly (p/n 5536-009.06),
mounted to the table bearing top-inner-frame assembly.
(4) Locate cam four on the top-drive control-switch assembly. Cam four is the
fourth cam from the gear reducer (p/n 6468.062.01).
(5)
Using a 1/4-inch wrench, slightly loosen the screw securing the position
of cam four.
Damage to the table can result if over-adjustment is done. Make small,
CAUTION:
incremental adjustments alternated with verification to ensure safe
operation.
(6)
Rotate cam four a small amount in either direction and secure the locking
screw.
(7)
Press the top-right switch to move the table top back toward center.
(8) Press top-left and observe the table top's new stop position in relation to
the 30-inch mark.
(9)
If the mark lines up with the table body end, adjustment is complete.
MD0351
1-19