Damage to the table can result if over-adjustment is done. Make small,
CAUTION:
incremental adjustments alternated with verification to ensure safe
operation.
(6)
Rotate cam two a small amount in either direction and secure the locking
screw.
(7)
Press "top-left" to move the table top to the left any amount.
(8)
Press "top-center" to center the table top.
(9)
If the table top ends align with the table body ends, adjustment is
complete.
(10) Further incremental adjustment is dependent upon the new stop
position.
(a) If the table traveled further out of alignment, cam two must be rotated
a small amount in the opposite direction. Go back to step (6) and proceed.
(b) If the table traveled closer to proper alignment, cam two is being
rotated in the proper direction. Go back to step (6) and proceed.
(11) Upon completion of adjustment, replace the table sub-top and the table
top.
o. Table Top Right-of-Center Position Adjustment.
(1)
Verify adjustment is necessary.
(a) Press top-center and ensure the table top centers itself by ensuring
the ends of the table-top frame align with the ends of the table body. If the ends are in
alignment, no adjustment is necessary.
(b) If the table top centers too far to the left (head) end of the table
body, repeat the steps in paragraph 1-4n.
(c) If the table top centers too far to the right (foot) end of the table
body, proceed to step (2).
MD0351
1-21